Made in the european union

The analyzer doesn’t work. On the display appears following message:

Error 10  or  Empty Sensor

            Possible reason and removal of the problem:

The sucked milk quantity is insufficient or there is air in the sample – please, check for air bubbles in the sample. Please check whether the milk sample reduces in the sample mug – if no, there is problem of  the device suction system.
 

The sample temperature is high

             Possible reason and removal of the problem:

Sucked sample is overheated. Milk sample temperature must be in the maximum permissible range.
.
 

The temperature of box is high

            Possible reason and removal of the problem:

Milk analyzer temperature exceeded the limits. Enable the option FAST START”(please see the SETTINGS in MENU mode – FAST START )
 
 

 Error 11

            Possible reason and removal of the problem:

Sensor is damaged – the sensor must be changed. Please, contact Milkotester for assistance.

Stay connected with our newsletter

Español