Error 10 or Empty Sensor
Possible reason and removal of the problem:
The sucked milk quantity is insufficient or there is air in the sample – please, check for air bubbles in the sample. Please check whether the milk sample reduces in the sample mug – if no, there is problem of the device suction system.
The sample temperature is high
Possible reason and removal of the problem:
Sucked sample is overheated. Milk sample temperature must be in the maximum permissible range.
.
The temperature of box is high
Possible reason and removal of the problem:
Milk analyzer temperature exceeded the limits. Enable the option FAST START”(please see the SETTINGS in MENU mode – FAST START )
Error 11
Possible reason and removal of the problem:
Sensor is damaged – the sensor must be changed. Please, contact Milkotester for assistance.